User ManUalRaySafe i2System
10RaySafe i2 System User Manual – Using the real time displayUSING THE REAL TIME DISPLAYINTRODUCTIONThe real time display is the display unit in the R
11RaySafe i2 System User Manual – Using the real time displayThe real time display screen is a touch screen. Operate the real time display by tapping
12RaySafe i2 System User Manual – Using the real time displayONLINE VIEW VIEW CURRENT DOSE RATES Figure 5. The online viewNOTE Dosimeters are sho
13RaySafe i2 System User Manual – Using the real time display6781 2 3 45Figure 6. The online view elementsLEGENd 1 Personal dose overview (see page
14RaySafe i2 System User Manual – Using the real time displayWhen a dosimeter moves out of range of the real time display, its button will remain visi
15RaySafe i2 System User Manual – Using the real time display• Dosimeter – a unique ID per dosimeter • Battery status • 2-4 blocks indicate normal
16RaySafe i2 System User Manual – Using the real time display• Dose session 1 and 2 values, date and time. Use the dose sessions to measure dose for
17RaySafe i2 System User Manual – Using the real time display• The annual dose limit distributed over the full year (dashed line). As long as the acc
18RaySafe i2 System User Manual – Using the real time displayIn the day/week time spans you can view accumulated dose value bars. Each bar represents
19RaySafe i2 System User Manual – Using the real time displayUSER SETTINGS MENU Figure 11. The user settings menuThe user settings menu consists of o
© 2013.01 Unfors RaySafe 5001047-CAll rights are reserved. Reproduction or transmission in whole or in part, in any form or by any means, electronic,
20RaySafe i2 System User Manual – Using the real time displayADMIN SETTINGS MENU The Admin Settings menu consists of four screens. Access these screen
21RaySafe i2 System User Manual – Using the real time displayReserved Slots menu Figure 13. The reserved slots menuThis is an optional setting, which
22RaySafe i2 System User Manual – Using the real time displayNetwork Setup menu Figure 14. The network setup menuConfigure the network connection betw
23RaySafe i2 System User Manual – Using the real time displayTime and Date menu Figure 15. The time and date menuSet time, date, time zone and manage
24RaySafe i2 System User Manual – MaintenanceMAINTENANCECALIBRATING THE REAL TIME DISPLAY SCREENIf the touch screen does not respond correctly to user
25RaySafe i2 System User Manual – MaintenanceChrome parts should only be cleaned by rubbing down with a dry cloth. Do not use abrasive polishes. To pr
26RaySafe i2 System User Manual – TroubleshootingTROUBLESHOOTINGREAL TIME DISPLAYTable 1. Real time display troubleshooting PRObLEm SOLUTION The dos
27RaySafe i2 System User Manual – TroubleshootingPRObLEm SOLUTION Dose viewer/dose manager does not detect the cradle • Check the USB connection b
28RaySafe i2 System User Manual – Important information About the systemIMPORTANT INFORMATION ABOUT THE SYSTEMINTENDED USE OF THE RAYSAFE I2 SYSTEMThi
29RaySafe i2 System User Manual – Important information About the systemor operation, may relieve the manufacturer (or his agent) from all or some res
3RaySafe i2 System User Manual – ContentsCONTENTSINTRODUCTION ...
30RaySafe i2 System User Manual – Important information About the systemSYSTEM REQUIREMENTSDose viewer and dose manager has the following system requi
31RaySafe i2 System User Manual – Technical dataTECHNICAL DATADOSIMETER RADIOLOGY ChARACTERISTIC mEASUREOperational Quantity* Hp(10)Reproducibility
32RaySafe i2 System User Manual – Technical data• The dose rate memory, where the dosimeter stores dose rate samples. When the radiation exceeds 40µ
33RaySafe i2 System User Manual – Technical dataFigure 16. Dose viewer/dose manager chart. In lack of dose rate samples, the yellow curve displays me
34RaySafe i2 System User Manual – Technical dataREAL TIME DISPLAY Table 3. Real time display technical specifications ChARACTERISTIC mEASURE Weight
35RaySafe i2 System User Manual – Technical dataChARACTERISTIC mEASURELog memory 2 3600 Dose Rate samples with one-second resolution, cyclically over
36RaySafe i2 System User Manual – Technical dataCLIMATIC CONDITIONSClasses: C1Indoors, temperature controlledPARAmETER mEASURETemperature levels:– for
37RaySafe i2 System User Manual – Technical dataSECURITY AND PRIVACY REQUIREMENTS ANTI VIRUS An AVS solution is not installed on this Windows CE devic
38RaySafe i2 System User Manual – SafetySAFETYIMPORTANT SAFETY DIRECTIONSIf the RaySafe i2 system is not functioning correct or damage is visible, ple
39RaySafe i2 System User Manual – SafetySAFETY DEVICES WARNING Never attempt to remove, modify, override or frustrate any safety device on the produc
4RaySafe i2 System User Manual – ContentsTECHNICAL DATA ...
40RaySafe i2 System User Manual – SafetyFIRE SAFETY Use of electrical product in an environment for which it was not designed can lead to fire or explo
41RaySafe i2 System User Manual – SafetyELECTROMAGNETIC COMPATIBILITY EMCThis Unfors RaySafe product complies with relevant international and nation
42RaySafe i2 System User Manual – Safetyagainst harmful interference when the equipment is operated in a commercial environment. This equipment genera
43RaySafe i2 System User Manual – SafetyNETWORK SAFETY, SECURITY AND PRIVACYCUSTOMER ROLE IN THE PRODUCT SECURITY PARTNERSHIP We recognize that the se
44RaySafe i2 System User Manual – SafetyEXPLANATION OF THE LABEL SYMBOLSSymbOL ExPLANATIONCaution, consult accompanying documentsConformity marking to
45RaySafe i2 System User Manual – SafetyREAL TIME DISPLAY PRODUCT LABEL The following real time display product label is located on the rear side of t
46RaySafe i2 System User Manual – SafetyIP21INPUT RATING: 12 V86 8,3 M HzModel: 8603024S/N: 0000 0000D ate of M fg :01 /20 1 01.2 ARaysafe i2 real t
47RaySafe i2 System User Manual – SafetyDate of Mfg:01/2010S/N: 0000 0000Model: 8601019 918,3MHzFCC ID:XWK8601015Electrical SafetyE113017IC: 9038A-8
48RaySafe i2 System User Manual – Product disposalPRODUCT DISPOSALINTRODUCTIONUnfors RaySafe is concerned to help protect the natural environment, and
49RaySafe i2 System User Manual – Product disposalIn the European Union (the WEEE directive), this label indicates that this product should not be dis
5RaySafe i2 System User Manual – IntroductionINTRODUCTIONABOUT THE RAYSAFE i2 SYSTEMFigure 1. RaySafe i2 System overview The RaySafe i2 System can co
6RaySafe i2 System User Manual – IntroductionHOW THE SYSTEM WORKSThe dosimeter measures and records X-ray exposure every second and transfers the data
7RaySafe i2 System User Manual – Using the DosimeterUSING THE DOSIMETERFigure 2. The dosimeterINTRODUCTIONThe dosimeter is an active dose meter desig
8RaySafe i2 System User Manual – Using the cradleUSING THE CRADLEFigure 3. The cradleINTRODUCTIONThe cradle is a dock station that lets you connect a
9RaySafe i2 System User Manual – Using dose viewer and dose managerUSING DOSE VIEWER AND DOSE MANAGERSee dose viewer and dose manager user manuals, fo