Raysafe i2 System Manual do Utilizador

Consulte online ou descarregue Manual do Utilizador para Equipamento Raysafe i2 System. RaySafe i2 System User Manual Manual do Utilizador

  • Descarregar
  • Adicionar aos meus manuais
  • Imprimir

Resumo do Conteúdo

Página 1 - RaySafe i2

User ManUalRaySafe i2System

Página 2

10RaySafe i2 System User Manual – Using the real time displayUSING THE REAL TIME DISPLAYINTRODUCTIONThe real time display is the display unit in the R

Página 3 - CONTENTS

11RaySafe i2 System User Manual – Using the real time displayThe real time display screen is a touch screen. Operate the real time display by tapping

Página 4

12RaySafe i2 System User Manual – Using the real time displayONLINE VIEW  VIEW CURRENT DOSE RATES Figure 5. The online viewNOTE Dosimeters are sho

Página 5 - INTRODUCTION

13RaySafe i2 System User Manual – Using the real time display6781 2 3 45Figure 6. The online view elementsLEGENd 1 Personal dose overview (see page

Página 6 - ABOUT THIS USER MANUAL

14RaySafe i2 System User Manual – Using the real time displayWhen a dosimeter moves out of range of the real time display, its button will remain visi

Página 7 - USING THE DOSIMETER

15RaySafe i2 System User Manual – Using the real time display• Dosimeter – a unique ID per dosimeter • Battery status • 2-4 blocks indicate normal

Página 8 - USING THE CRADLE

16RaySafe i2 System User Manual – Using the real time display• Dose session 1 and 2 values, date and time. Use the dose sessions to measure dose for

Página 9

17RaySafe i2 System User Manual – Using the real time display• The annual dose limit distributed over the full year (dashed line). As long as the acc

Página 10 - USING THE REAL TIME DISPLAY

18RaySafe i2 System User Manual – Using the real time displayIn the day/week time spans you can view accumulated dose value bars. Each bar represents

Página 11 - REAL TIME DISPLAY FEATURES

19RaySafe i2 System User Manual – Using the real time displayUSER SETTINGS MENU Figure 11. The user settings menuThe user settings menu consists of o

Página 12

© 2013.01 Unfors RaySafe 5001047-CAll rights are reserved. Reproduction or transmission in whole or in part, in any form or by any means, electronic,

Página 13

20RaySafe i2 System User Manual – Using the real time displayADMIN SETTINGS MENU The Admin Settings menu consists of four screens. Access these screen

Página 14

21RaySafe i2 System User Manual – Using the real time displayReserved Slots menu Figure 13. The reserved slots menuThis is an optional setting, which

Página 15

22RaySafe i2 System User Manual – Using the real time displayNetwork Setup menu Figure 14. The network setup menuConfigure the network connection betw

Página 16

23RaySafe i2 System User Manual – Using the real time displayTime and Date menu Figure 15. The time and date menuSet time, date, time zone and manage

Página 17

24RaySafe i2 System User Manual – MaintenanceMAINTENANCECALIBRATING THE REAL TIME DISPLAY SCREENIf the touch screen does not respond correctly to user

Página 18

25RaySafe i2 System User Manual – MaintenanceChrome parts should only be cleaned by rubbing down with a dry cloth. Do not use abrasive polishes. To pr

Página 19

26RaySafe i2 System User Manual – TroubleshootingTROUBLESHOOTINGREAL TIME DISPLAYTable 1. Real time display troubleshooting PRObLEm SOLUTION The dos

Página 20

27RaySafe i2 System User Manual – TroubleshootingPRObLEm SOLUTION Dose viewer/dose manager does not detect the cradle • Check the USB connection b

Página 21

28RaySafe i2 System User Manual – Important information About the systemIMPORTANT INFORMATION ABOUT THE SYSTEMINTENDED USE OF THE RAYSAFE I2 SYSTEMThi

Página 22

29RaySafe i2 System User Manual – Important information About the systemor operation, may relieve the manufacturer (or his agent) from all or some res

Página 23

3RaySafe i2 System User Manual – ContentsCONTENTSINTRODUCTION ...

Página 24 - MAINTENANCE

30RaySafe i2 System User Manual – Important information About the systemSYSTEM REQUIREMENTSDose viewer and dose manager has the following system requi

Página 25 - DISINFECTION

31RaySafe i2 System User Manual – Technical dataTECHNICAL DATADOSIMETER RADIOLOGY ChARACTERISTIC mEASUREOperational Quantity* Hp(10)Reproducibility

Página 26 - TROUBLESHOOTING

32RaySafe i2 System User Manual – Technical data• The dose rate memory, where the dosimeter stores dose rate samples. When the radiation exceeds 40µ

Página 27 - OTHER USER MANUALS

33RaySafe i2 System User Manual – Technical dataFigure 16. Dose viewer/dose manager chart. In lack of dose rate samples, the yellow curve displays me

Página 28

34RaySafe i2 System User Manual – Technical dataREAL TIME DISPLAY Table 3. Real time display technical specifications ChARACTERISTIC mEASURE Weight

Página 29 - TECHNICAL DOCUMENTATION

35RaySafe i2 System User Manual – Technical dataChARACTERISTIC mEASURELog memory 2 3600 Dose Rate samples with one-second resolution, cyclically over

Página 30 - SYSTEM REQUIREMENTS

36RaySafe i2 System User Manual – Technical dataCLIMATIC CONDITIONSClasses: C1Indoors, temperature controlledPARAmETER mEASURETemperature levels:– for

Página 31 - TECHNICAL DATA

37RaySafe i2 System User Manual – Technical dataSECURITY AND PRIVACY REQUIREMENTS ANTI VIRUS An AVS solution is not installed on this Windows CE devic

Página 32 - REAL TIME DISPLAY MEMORY

38RaySafe i2 System User Manual – SafetySAFETYIMPORTANT SAFETY DIRECTIONSIf the RaySafe i2 system is not functioning correct or damage is visible, ple

Página 33 - TECHNICAL SPECIFICATIONS

39RaySafe i2 System User Manual – SafetySAFETY DEVICES WARNING Never attempt to remove, modify, override or frustrate any safety device on the produc

Página 34 - DOSIMETER

4RaySafe i2 System User Manual – ContentsTECHNICAL DATA ...

Página 35

40RaySafe i2 System User Manual – SafetyFIRE SAFETY Use of electrical product in an environment for which it was not designed can lead to fire or explo

Página 36 - CLIMATIC CONDITIONS

41RaySafe i2 System User Manual – SafetyELECTROMAGNETIC COMPATIBILITY EMCThis Unfors RaySafe product complies with relevant international and nation

Página 37 - ENCRYPTION

42RaySafe i2 System User Manual – Safetyagainst harmful interference when the equipment is operated in a commercial environment. This equipment genera

Página 38 - IMPORTANT SAFETY DIRECTIONS

43RaySafe i2 System User Manual – SafetyNETWORK SAFETY, SECURITY AND PRIVACYCUSTOMER ROLE IN THE PRODUCT SECURITY PARTNERSHIP We recognize that the se

Página 39 - EXPLOSION SAFETY

44RaySafe i2 System User Manual – SafetyEXPLANATION OF THE LABEL SYMBOLSSymbOL ExPLANATIONCaution, consult accompanying documentsConformity marking to

Página 40 - ELECTROSTATIC DISCHARGE ESD

45RaySafe i2 System User Manual – SafetyREAL TIME DISPLAY PRODUCT LABEL The following real time display product label is located on the rear side of t

Página 41

46RaySafe i2 System User Manual – SafetyIP21INPUT RATING: 12 V86 8,3 M HzModel: 8603024S/N: 0000 0000D ate of M fg :01 /20 1 01.2 ARaysafe i2 real t

Página 42 - MODALITY SPECIFIC SAFETY

47RaySafe i2 System User Manual – SafetyDate of Mfg:01/2010S/N: 0000 0000Model: 8601019 918,3MHzFCC ID:XWK8601015Electrical SafetyE113017IC: 9038A-8

Página 43 - EQUIPMENT LABEL OVERVIEW

48RaySafe i2 System User Manual – Product disposalPRODUCT DISPOSALINTRODUCTIONUnfors RaySafe is concerned to help protect the natural environment, and

Página 44

49RaySafe i2 System User Manual – Product disposalIn the European Union (the WEEE directive), this label indicates that this product should not be dis

Página 45 - S/N: 0000 0000

5RaySafe i2 System User Manual – IntroductionINTRODUCTIONABOUT THE RAYSAFE i2 SYSTEMFigure 1. RaySafe i2 System overview The RaySafe i2 System can co

Página 46 - DOSIMETER PRODUCT LABEL

6RaySafe i2 System User Manual – IntroductionHOW THE SYSTEM WORKSThe dosimeter measures and records X-ray exposure every second and transfers the data

Página 47

7RaySafe i2 System User Manual – Using the DosimeterUSING THE DOSIMETERFigure 2. The dosimeterINTRODUCTIONThe dosimeter is an active dose meter desig

Página 48 - PRODUCT DISPOSAL

8RaySafe i2 System User Manual – Using the cradleUSING THE CRADLEFigure 3. The cradleINTRODUCTIONThe cradle is a dock station that lets you connect a

Página 49

9RaySafe i2 System User Manual – Using dose viewer and dose managerUSING DOSE VIEWER AND DOSE MANAGERSee dose viewer and dose manager user manuals, fo

Comentários a estes Manuais

Sem comentários